Airon pNeuton Model A Ventilator

Portable Ventilators

For Rent or Purchase

Condition: New

REQUEST QUOTE

The pNeuton model A ventilator with built-in CPAP (Continuous Positive Airway Pressure) is ideal for critical care adult or pediatric patient support in hospital environments.

With simple-to-use controls – Tidal Volume, Respiratory Rate, and Pressure Limit – pNeuton A can provide volume-targeted or pressure-limited ventilation. The built-in high output demand flow CPAP system insures critical care ventilation needs are optimized for spontaneous and mandatory breaths.

The pNeuton A is fully pneumatic and operational without the need of batteries or electrical power, helping you care for the most difficult patients, wherever they may be located. It has an integrated all-pneumatic alarm system with remote alarm capability that keeps you alerted to patient circuit disconnects and low gas supply.

Patient Benefits

  • Provide invasive or non-invasive patient support
  • Low patient Work of Breathing with peak flow of up to 140 L/min to meet patient needs
  • 100% or 65% to match patient oxygen requirements
  • Switch from mask CPAP to invasive ventilation seamlessly and without interruption
  • Low-profile, easy-seal mask reduces facial pressure

Clinician Benefits

  • Calibrated CPAP and oxygen controls for easy operation
  • MRI conditional up to 3 T – place right next to the scanner with remote alarm output for the control room
  • Built-in CPAP/PEEP system with automatic sensitivity for changing patient needs; no need for external CPAP valves
  • Manometer to monitor patient pressure and effort
  • Monitor CO2 with invasive/non-invasive ventilation
  • Use the Safe-T-Vent Medication Nebulizer System and fill the nebulizer without losing CPAP
  • Use the bacteria / viral filter to protect caregiver from patient exhilation
  • Built-in patient disconnect alarm with remote output to nurse call systems
  • Oxygen powered – no batteries to change
  • Rugged and lightweight (6.5 lbs / 3 kg)

Benefits for your Budget

  • Two oxygen levels; conserve oxygen with 65% setting
  • 60-minute CPAP run time on a “E” cylinder
  • Demand flow CPAP system with minimal wasted oxygen
  • Standardized economical patient circuits save on operating costs
  • No batteries or electronics decrease maintenance cost

Specs

pNeuton Model A P/N 20052

  • Pneumatically operated for use on pediatric and adult patients > 23 kg
  • Demand flow system for spontaneous breathing to 140 L/min
  • Pressures displayed on manometer
  • Modes of Operation: CMV, IMV, CPAP, Pressure Limit
  • Mandatory breath flow rate is fixed at 36 L/min
  • MRI compatible: up to 3 T static field strength, 6.9 G/cm spatial field gradient, 300 V RF power, shielded, no gauss line restriction
  • Weight: 6.5 pounds (3 kg)
  • Dimensions: 5″H x 10″W x 7″D (13 x 25 x 18 cm)
  • Input gas requirement: 55 psi ± 15 psi (3.8 bar ± 1 bar)
  • Classification according to MDD 93/42/EEC: IIb
  • Meets International Standards for transport ventilators:
    • ASTM: Ventilators Intended for Use in Critical Care (F 1100-90)
    • ISO: Lung Ventilators for Medical Use, Part 3 (EN 794-3:1999)

Control Settings

Mandatory Breaths: ON or OFF
Respiratory Rate: 3 to 50 bpm
Tidal Volume: 360 to 1,500 ml
Peak Pressure: 15 to 75 cm H2O (mbar)
PEEP/CPAP: 0 to 20 cm H2O (mbar)
Oxygen: 100% or 65%

Audible and Visual Alarms

  • All pneumatic alarm system (no batteries – requires minimum 30 psi / 2 bar gas supply pressure)
  • Patient circuit disconnection
  • Automatic reset when alarm condition resolves
  • 22 Second Activation
  • One minute silence / reset button
  • Low oxygen inlet pressure (below 40 psi)
  • Remote alarm output connection