The pNeuton model S ventilator provides mask CPAP plus volume or pressure ventilation for pediatric to adult patients for pre-hospital and hospital environments; with simple-to-use controls – Tidal Volume, Respiratory Rate and Pressure Limit.
For early intervention, the built-in CPAP (Continuous Positive Airway Pressure) system allows you to start with face mask CPAP and if the patient needs it, seamlessly provide ventilation without interruption. Fully pneumatic, the pNeuton S does not use batteries or electrical power to provide ventilation with readiness for operation.
The built-in high output demand flow CPAP system with automatic sensitivity insures critical care ventilation needs are optimized for spontaneous and mandatory breaths.
Patient Benefits
- Provides invasive or non-invasive patient support
- Low patient Work of Breathing with peak flow of up to 140 L/min to meet patient needs
- 100% or 65% to match patient oxygen requirements
- Switch from mask CPAP to invasive ventilation seamlessly and without interruption
- Low-profile, easy-seal mask reduces facial pressure
Clinician Benefits
- Calibrated CPAP and oxygen controls for easy operation
- MRI conditional up to 3 T – place right next to the scanner with remote alarm output for the control room
- Manometer to monitor patient pressure and effort
- Built-in CPAP / PEEP system with automatic sensitivity for changing patient needs; no need for external CPAP valves
- Night visible pressure gauge for patient monitoring
- Monitor CO2 with invasive/non-invasive ventilation
- Oxygen powered – no batteries to change
- Rugged and lightweight (6 lbs / 2.7 kg)
- Use the bacteria / viral filter to protect caregiver from patient exhalation
Benefits for Your Budget
- Two oxygen levels; conserve oxygen with 65% setting
- 45-minute CPAP run time on a D cylinder
- Demand flow CPAP system with minimal wasted oxygen
- Economical patient circuits save on operating costs
- No batteries or electronics decrease maintenance cost
Specs
pNeuton Model S P/N 20002
- Pneumatically operated for use on pediatric and adult patients > 23 kg
- Demand flow system for spontaneous breathing to 140 L/min
- Pressures displayed on manometer
- Modes of Operation: CMV, IMV, CPAP, Pressure Limit
- Mandatory breath flow rate is fixed at 36 L/min
- MRI conditional – 3 T or less, maximum spacial gradient magnetic field of 720-gauss/cm or less, no gauss line restriction
- Weight: 6 pounds (2.7 kg)
- Dimensions: 4″ H x 8″ W x 6″ D (10 x 20 x 15 cm)
- Input gas requirement: 55 psi ± 15 psi (3.8 bar ± 1 bar)
- Classification according to MDD 93/42/EEC: IIb
- Meets International Standards for transport ventilators:
- ASTM: Ventilators Intended for Use in Critical Care (F 1100-90)
- ISO: Lung Ventilators for Medical Use, Part 3 (EN 794-3:1999)
Control Settings
Mandatory Breaths: ON or OFF
Respiratory Rate: 2 to 50 bpm
Tidal Volume: 360 to 1,500 ml
Peak Pressure: 10 to 75 cm H2O (mbar)
PEEP / CPAP: 0 to 20 cm H2O (mbar)
Oxygen: 100% or 65%
Audible and Visual Alarms
- All pneumatic alarm system (no batteries – requires minimum 30 psi / 2 bar gas supply pressure)
- Low gas source pressure
- If source gas drops below 40 psi (2 bar)
- Continues operation as long as oxygen is available