Trilogy EVO Medical Device Correction

Trilogy EVO Medical Device Correction

This is a Philips URGENT Medical Device Correction letter for the Trilogy Evo and Trilogy EV300 Urgent Field Safety Notice that could pose a risk for patients or users. This letter is intended to inform you about:

  • what the problems are and under what circumstances they can occur,
  • the actions that should be taken by the customer/user in order to prevent risks for patients, and
  • the actions planned by Philips to correct the problem.

Important Points from USME

  • Devices can remain in use
  • The software fix will come in a separate communication and field action
  • In the EFS Calibration issue (issue 1), the device will alarm as designed. This issue has no reported customer complaints/incidents and was discovered during our follow testing on the Pressure Drift issue only.
  • Pressure drift (issue 2) is associated with long-term use (2cmH2O over a month of uninterrupted use) High pressure alarms will annunciate when pressure threshold is reached.
Trilogy EVO Medical Device Correction

1. What the problem is and under what circumstances it can occur

Two software issues have been identified related to pressure increase. The first issue is described as Infant/Pediatric Infant/Pediatric EFS Calibration Pressure Increase and the second issue is described as Pressure Drift (Continuous Usage).

Issue 1 – Infant/Pediatric EFS Calibration Pressure Increase

An increase in the expiratory pressure (EPAP/PEEP) can occur when the pediatric/infant External Flow Sensor (EFS) is used with an Active Flow or Dual Limb circuit and a manual circuit calibration is performed. This increase in pressure will be seen shortly after starting therapy within approximately 1 to 2 minutes. The maximum expiratory pressure increase may reach up to 10 cmH2O above the set pressure. The inspiratory pressure (Pressure Support/Pressure Control/IPAP) may also be affected.

Issue 2 – Pressure Drift (Continuous Usage)

When a Trilogy Evo is used continuously without any interruption of therapy over weeks to months, the baseline pressure (that is, the pressure initially set for the patient) may increase or decrease at a rate of up to approximately 2 cmH2O per month. This increase applies to the PEEP and the inspiratory pressure at the same rate. The maximum pressure deviation that could be seen is a 10 cmH2O shift from the baseline pressure. The pressure regulation alarms will not annunciate with this issue. This issue applies to the following modes: CPAP, PSV, S/T, A/C-PC, SIMV-PC, SIMV-VC (PEEP and Pressure Support), and A/C-VC (PEEP). This document contains important information for the continued safe and proper use of your equipment.

Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication.

The User Interface (display screen) or Care Orchestrator/Care Orchestrator Essence display the actual pressure that the patient is receiving, which will differ from the baseline setting when this condition occurs.

2. Describe the hazard/harm associated with the issue

Issue 1 – Infant/Pediatric EFS Calibration Pressure Increase

There is potential for inappropriate therapy to be provided when the EFS is used with manual calibration. Harms that could occur as a result of an increase in delivered pressure include: Barotrauma (an injury to the lungs because of too much pressure) Hypotension (a decrease in blood pressure) If the expiratory pressure (EPAP/PEEP) increases but the inspiratory pressure (Pressure Support/Pressure Control/IPAP) remains unchanged, the following harm could occur: Hypercarbia (too much carbon dioxide in the blood)

Issue 2 – Pressure Drift (Continuous Usage)

There is potential for inappropriate therapy to be provided to a patient if there is a pressure increase or decrease from the baseline pressure settings without alarm. Harms that could occur as a result of pressure drift increase over time include:

  • Barotrauma (an injury to the lungs as a result of too much pressure). The potential occurrence of barotrauma related to this issue is considered to unlikely; however, if it were to occur, it may be a serious injury.
  • Hypotension (a decrease in blood pressure). The potential occurrence of hypotension related to this issue is considered to be occasional.

Harms that could occur as a result of pressure drift increase over time, and then abruptly reverting therapy (pressure settings) back to the initial prescribed settings, include: 

  • Dyspnea (a sensation of shortness of breath)
  • Hypoxemia (a low amount of oxygen in the blood)

Harms that could occur as a result of pressure drift decrease over time include:

  • Hypoxemia (a low amount of oxygen in the blood)

3. Affected products and how to identify them

Product name, Product number

Trilogy Evo: BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B,

ES2110X15B, EU2110X15B, FR2110X14B, GB2110X15B, IA2110X15B,

IN2110X15B, IT2110X21B, KR2110X15B, LA2110X15B, LD2110X23B,

ND2110X15B, RDE2110X13B, UDS2110X11B

Trilogy Evo devices with software versions 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06,

1.05.01 and 1.06.02 are impacted. 

To check the software version, tap the Options icon located on the top left side of the screen. Choose “Information” to find the software version.

4. What you should do

Issue 1 – Infant/Pediatric EFS Calibration Pressure Increase

Please take the following actions until the upcoming software fix is implemented on your device. If you are using the infant/pediatric EFS, do not perform a manual circuit calibration. Instead, use the default calibration. The pressure increase will not occur if the default circuit calibration is used. As stated in the Clinical Manual, the ventilator is optimized for circuits that are within the specifications shown in “Circuit Requirements” section of the Clinical Manual, which are listed below. When using the default circuit calibration settings, ensure that you are using a circuit with these specifications:

  • Inspiratory/expiratory resistance: up to 5 cmH2O at:

o 15 L/min for pediatric (14 to 16 mm)

o 2.5 L/min for infant (9 to 13 mm) circuit size

  • Compliance: up to 4 ml/cmH2O

If your circuit does not meet these criteria, then seek other options as stated in the Clinical Manual. When setting volumes that are greater than or equal to 50 ml, passive and active PAP circuit types may be used. If, the above options are not clinically appropriate, then seek an alternative ventilator.

Issue 2 – Pressure Drift (Continuous Usage)

The following methods should be used to inspect the device to determine if a pressure drift has occurred as the pressure alarms will not detect it:

i. The Respiratory clinician or Physician can detect the increase/decrease in pressure by observing the pressure Trend report in Care Orchestrator/Care Orchestrator Essence. If a pressure drift is detected, the decision to reset the device and return to its original, intended settings MUST be made in consultation with a clinician. Take steps to support the patient during a brief interruption in therapy, as needed.

5. What Philips is Doing to Resolve this Recall

Philips is releasing this Urgent Medical Device Recall Letter to make customers and consignees aware of the potential issues described above and how to address them. Philips will be releasing a software correction to be made available via the “My Philips for Professionals” website for customers to upgrade devices. Philips will be contacting Trilogy Evo customers when the software is released.

Should you have any questions, please reach out to the US Med-Equip Quality Assurance Team at QA@usmedequip.com.

If you need any further information or support concerning this issue, please contact Philips Customer Care Solutions Center +1-800-345-6443 Option 2, Option 1 (Available 24/7) This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA’s MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax.

Philips regrets any inconvenience caused by this problem.

About US Med-Equip

US Med-Equip (USME) partners with top hospitals across the nation to provide the highest quality movable medical equipment for patients in their care. USME, an Inc. 5000 Fastest Growing company, supplies quality-certified equipment rented, sold, maintained and managed using the latest technology to help healthcare providers focus on their patients’ healing. Contact USME today for your critical care equipment rental needs, medical equipment rentals, or medical equipment for sale.

 